Manu
PRAKASH, et al.
Foldscope
: Paper Microscope
http://foldscope.com
Foldscope:
Microscopy for everyone
We are a research team at PrakashLab at Stanford University,
focused on democratizing science by developing scientific tools
that can scale up to match problems in global health and science
education. Here we describe Foldscope, a new approach for mass
manufacturing of optical microscopes that are printed-and-folded
from a single flat sheet of paper, akin to Origami.
http://scopeblog.stanford.edu/2014/03/10/stanford-bioengineer-develops-a-50-cent-paper-microscope/
March 10th, 2014
Stanford
bioengineer develops a 50-cent paper microscope
by
Kris
Newby
Stanford bioengineer develops a 50-cent paper microscope
UPDATE: A second blog entry, including a link to Prakash’s TED
talk on this topic, can be found here. And this entry discusses
Prakash’s plans to give away 10,000 build-your-own paper
microscope kits to citizen scientists with the most inspiring
ideas for things to do with this new invention.
When Manu Prakash, PhD, wants to impress lab visitors with the
durability of his Origami-based paper microscope, he throws it
off a three-story balcony, stomps on it with his foot and dunks
it into a water-filled beaker. Miraculously, it still works.
Even more amazing is that this microscope — a bookmark-sized
piece of layered cardstock with a micro-lens — only costs about
50 cents in materials to make.
In the video posted above, you can see his “Foldscope” being
built in just a few minutes, then used to project giant images
of plant tissue on the wall of a dark room.
Prakash’s dream is that this ultra-low-cost microscope will
someday be distributed widely to detect dangerous blood-borne
diseases like malaria, African sleeping sickness,
schistosomiasis and Chagas.
“I wanted to make the best possible disease-detection instrument
that we could almost distribute for free,” said Prakash. “What
came out of this project is what we call use-and-throw
microscopy.”
The Foldscope can be assembled in minutes, includes no
mechanical moving parts, packs in a flat configuration, is
extremely rugged and can be incinerated after use to safely
dispose of infectious biological samples. With minor design
modifications, it can be used for bright-field,
multi-fluorescence or projection microscopy.
One of the unique design features of the microscope is the use
of inexpensive spherical lenses rather than the precision-ground
curved glass lenses used in traditional microscopes. These
poppy-seed-sized lenses were originally mass produced in various
sizes as an abrasive grit that was thrown into industrial
tumblers to knock the rough edges off metal parts. In the
simplest configuration of the Foldscope, one 17-cent lens is
press-fit into a small hole in the center of the slide-mounting
platform. Some of his more sophisticated versions use multiple
lenses and filters.
To use a Foldscope, a sample is mounted on a microscope slide
and wedged between the paper layers of the microscope. With a
thumb and forefinger grasping each end of the layered paper
strip, a user holds the micro-lens close enough to one eye that
eyebrows touch the paper. Focusing and locating a target object
are achieved by flexing and sliding the paper platform with the
thumb and fingers.
Because of the unique optical physics of a spherical lens held
close to the eye, samples can be magnified up to 2,000 times.
(To the right are two disease-causing microbes, Giardia lamblia
and Leishmania donovani, photographed through a Foldscope.)
The Foldscope can be customized for the detection of specific
organisms by adding various combinations of colored LED lights
powered by a watch battery, sample stains and fluorescent
filters. It can also be configured to project images on the wall
of a dark room.
In addition, Prakash is passionate about mass-producing the
Foldscope for educational purposes, to inspire children — our
future scientists — to explore and learn from the microscopic
world.
In a recent Stanford bioengineering course, Prakash used the
Foldscope to teach students about the physics of microscopy. He
had the entire class build their own Foldscope. Then teams wrote
reports on microscopic observations or designed Foldscope
accessories, such a smartphone camera attachment.
For more on Foldscope optics, a materials list and construction
details, read Prakash’s technical paper :
http://arxiv.org/ftp/arxiv/papers/1403/1403.1211.pdf
http://arxiv.org/abs/1403.1211
Foldscope:
Origami-based paper microscope
James Cybulski, James Clements, Manu Prakash
Here we describe an ultra-low-cost origami-based approach for
large-scale manufacturing of microscopes, specifically
demonstrating brightfield, darkfield, and fluorescence
microscopes. Merging principles of optical design with origami
enables high-volume fabrication of microscopes from 2D media.
Flexure mechanisms created via folding enable a flat compact
design. Structural loops in folded paper provide kinematic
constraints as a means for passive self-alignment. This light,
rugged instrument can survive harsh field conditions while
providing a diversity of imaging capabilities, thus serving
wide-ranging applications for cost-effective, portable
microscopes in science and education.
http://www.youtube.com/watch?v=pBjIYB5Yk2I
Foldscope
Print your own 50-cent microscope
This bookmark-sized microscope can be assembled in minutes,
includes no mechanical moving parts, packs in a flat
configuration, is extremely rugged and can be incinerated after
to safely dispose of infectious biological samples. With minor
optics modifications, the microscope can be designed for
brightfield, multi-flourescence or projection microscopy, or
specialized to identify specific pathogens.
http://arxiv.org/abs/1403.1211
http://en.wikipedia.org/wiki/Foldscope
Foldscope
A Foldscope is an optical microscope that can be assembled from
simple components, including a sheet of paper and a lens. It was
developed by Manu Prakash and designed to cost less than US$1 to
build.
Overview
A Foldscope is an optical microscope that can be assembled from
a punched sheet of cardstock, a spherical lens, a light emitting
diode and a diffuser panel, along with a watch battery that
powers the LED.[1] It can magnify up to 2000 times and weighs 8
grams. The magnification power is enough to enable the spotting
of organisms such as Leishmania donovani and Escherichia coli,
as well as malarial parasites.[2] A Foldscope can be printed on
a standard A4 sheet of paper and assembled in seven minutes.[3]
The basic principle of using a small spherical lens held close
to the eye dates back to Antonie van Leeuwenhoek (1632-1723),
who was the first to see single-celled organisms using such a
lens held in a device of his own design.[citation needed]
The Foldscope was developed by a team led by Manu Prakash, an
assistant professor of bioengineering at the Stanford School of
Medicine.[4] The project was funded by several organisations
including the Bill & Melinda Gates Foundation, which gave a
grant of US $100,000 for research in November 2012. The Gordon
and Betty Moore Foundation funded the "Ten Thousand Microscopes"
project under which Prakash plans to give away 10,000 Foldscope
kits to interested parties, including students for
research.[5][2][6] Foldscope sets will also be produced and
tested in Kenya, India and Uganda.[3]
Twelve Foldscope variants are available, each designed to aid
the identification of a particular disease-causing organism.[3]
To enable several people to use them at once, each microscope
can project images with a built-in projector.[3] The Foldscope
is designed to be assembled by the end user, and hence is
colour-coded to help with the assembly.[2] Each unit costs less
than one US dollar to build, with estimates varying from 50
cents to 97 cents.[2][3]
Foldscope
: Origami-Based Paper Microscope
by
James Cybulski, et al.
[ PDF, 2.4
MB ]
http://www.ted.com/talks/manu_prakash_a_50_cent_microscope_that_folds_like_origami
A
50-cent microscope that folds like origami
TEDGlobal 2012 · 9:21 · Filmed Jun 2012
Perhaps you’ve punched out a paper doll or folded an origami
swan? TED Fellow Manu Prakash and his team have created a
microscope made of paper that's just as easy to fold and use. A
sparkling demo that shows how this invention could revolutionize
healthcare in developing countries … and turn almost anything
into a fun, hands-on science experiment.
WO2013120091
OPTICAL
DEVICE
Inventor(s): PRAKASH MANU; CYBULSKI JAMES;
CLEMENTS JAMES
An optical device, such as a microscope, is disclosed that can
be assembled from flat materials. The optical device can be
assembled via a series of folds of a flat material. The optical
microscope can include a stage for supporting a sample, an optic
stage, and a light source. The optic stage can include one or
more lenses. The optical microscope can be capable of obtaining
simultaneous images from different forms of microscopy. The
optical microscope may have bright field and filter field
viewing capabilities wherein a user shifts from bright field to
filter field by lateral movement of the stage containing a lens
and a light source that cooperate to provide either the bright
field or the filter field.
FIELD
[0004] The present invention relates generally to diagnostic
optical devices, such as microscopes.
BACKGROUND
[0005] An optical instrument is an instrument used to move light
along a specified path or paths. Microscopes are common,
general-purpose optical instruments. Other optical instruments
include interferometers and spectrophotometers.
[0006] Microscopes are generally used to view objects that are
too small to be seen by the unaided eye. Optical microscopes use
visible light and a system of lenses to magnify images of small
objects. Optical microscopes are used in observing small
structures, determining pathology and diagnosing disease.
[0007] There are two basic configurations of the conventional
optical microscope, the simple (one lens) and compound (many
lenses). There are also several different types of microscopy
including brightfield, darkfield, fluorescence, and other forms.
Each of these forms of microscopy are performed individually and
one at a time. High magnification optical microscopes are often
heavy and take up much volume.
[0008] There is a need for improved optical instruments to
better diagnose pathologies and disease.
SUMMARY OF
THE DISCLOSURE
[0009] In some embodiments an optical device is provided. The
optical devices can include a first stage for supporting a
sample and a second stage engaged with the first stage and
movable relative to the first stage, the second stage including
an optic, the optic having a distance of less than about 3 mm
from the sample to the opposite side of the optic. In some
embodiments the optical device is an optical microscope. In some
embodiments the optical device is an interferometer or
spectrophotometers. In some embodiments the second stage can be
movable laterally relative to the first stage such that the
optic can be positioned over a desired location on the sample.
[00010] In some embodiments the first stage is configured to
receive a substrate containing the sample. In some embodiments
the first stage includes a slot shaped to receive the substrate.
In some embodiments the substrate is a glass slide. In some
embodiments the glass slide includes a portion treated with a
reagent to interact with the sample. In some embodiments the
sample is supported directly on the first stage. In some
embodiments the first stage includes a sample area comprising a
reagent to react with the sample.
[00011] In some embodiments the optical device includes an
illumination stage comprising a light source. The illumination
stage can be engaged with the first stage and second stage with
the illumination stage movable with the second stage. The light
source can be positioned adjacent to the first stage to
facilitate viewing the sample using the optic.
[00012] In some embodiments the light source is on one side of
the first stage and the optic is spaced apart from the opposite
side of the first stage a distance between the light source and
the side of the optic farthest from the first stage is less than
about 5 mm. In some embodiments the light source is on one side
of the first stage and the optic is spaced apart from the
opposite side of the first stage a distance between the light
source and the side of the optic farthest from the first stage
is from about 1 mm to about 20 mm.
[00013] In some embodiments the light source includes one or
more light-emitting diodes (LEDs). In some embodiments the LED
is white to provide a full-spectrum color image of the sample in
bright-field. In some embodiments the LED is blue to provide
light of an appropriate wavelength for fluorescence imaging. In
some embodiments the LED has an output power suitable to project
an image of the sample. In some embodiments the LED has a power
suitable to illuminate the sample so that a user's eye may
perceive the image of the sample. In some embodiments the LEDs
have power suitable so that a user may perceive multiple
simultaneous images of the sample.
[00014] In some embodiments the optical device includes an
illumination stage with an aperture. In some embodiments the
illumination stage includes an aperture having a diameter of
about 1/4 to about 2/3 of the diameter of the substantially
spherical lens. In some embodiments the optical microscope
includes an illumination stage with an element for modifying the
profile of the light from the LED, such as a condenser,
diffuser, light shaping diffuser, polycarbonate light shaping
filter, etc. In some embodiments, optical microscope has an
illumination stage that provides Kohler illumination.
[00015] In some embodiments the optical device is manufactured
from a flat material that includes the first stage and second
stage. In some embodiments a single material piece provides the
first stage and second stage. In some embodiments the flat
material comprises one or more of paper, polymer, and metal.
[00016] In some embodiments a power source is provided with the
optical microscope. The power source can engage with the
illumination stage and is configured to provide power to the
light source.
[00017] In some embodiments the microscope optic comprises a
substantially spherical ball lens. In some embodiments the
substantially spherical ball lens has a diameter of less than
about 2,500 [mu][iota][eta]. In some embodiments the spherical
ball lens has a diameter of about 1 ,000 [mu][iota][eta] to
about 2,500 [mu][pi][iota]. In some embodiments the spherical
ball lens has a diameter of about 300 [mu][eta][iota] to about 1
,000 [mu][pi][iota]. In some embodiments the spherical ball lens
has a diameter of about 100 [mu][iota][eta] to about 300
[mu][eta][iota]. In some embodiments the spherical lens has a
diameter of about 200 [mu][iota][eta] to about 1,000
[mu][pi][iota].
[00018] In some embodiments the spherical lens has an effective
aperture of less than the diameter of the spherical lens. In
some embodiments the aperture is about 1/4 to about 2/3 of the
diameter of the spherical lens. In some embodiments the
spherical ball lens has an aperture diameter of about 1/4 to 1/3
the lens diameter. In some embodiments the spherical lens has an
aperture diameter of about 1/3 to 1/2 the lens diameter. In some
embodiments the spherical lens has an aperture diameter of about
1/2 to 2/3 the lens diameter. In some embodiments half-ball
spherical lenses can be used. In some embodiments a Wallston
doublet lens is utilized. The Wallston doublet lens can be
composed of multiple half ball lenses. In some embodiments a
Gradient Index lens is used.
[00019] In some embodiments the optical microscope is
manufactured from a flat material that includes the first stage,
second stage, and illumination stage. In some embodiments, the
optical microscope is manufactured in a single instance. In some
embodiments, a series of folds produce the final configuration
of the microscope. The optical microscope can have an optical
alignment of the illumination stage, first stage, and second
stage achieved passively by separating and folding the flat
material. The optical alignment can have an accuracy of about 10
microns or less.
[00020] In some embodiments, the folding accuracy is
accomplished by geometrical features cut in flat material that
act as kinematic couplings thus providing a self-alignment. In
some embodiments, self-alignment is further improved by
providing structural closed loops in folding steps.
[00021] In some embodiments, the optical microscope can have an
integrated microfluidic channel for bringing samples directly to
the microscope lens.
[00022] In some embodiments, the optical microscope can be
incinerated after one use safely and thus can be used with
infected samples.
[00023] In some embodiments, the entire microscope is disposable
after single or multiple uses.
[00024] In some embodiment, a waveguide is utilized to channel
light from the light source to other optical components.
[00025] In some embodiments the second stage of the optical
microscope includes an opening with the lens in the opening.
[00026] In some embodiments the microscope has a magnification
of about 100X to 200X. In some embodiments the microscope has a
magnification of about 200X to 500X. In some embodiments the
microscope has a magnification of about 500X to 1,500X. In some
embodiments the microscope has a magnification of about 1 ,500X
to 2,500X. In some embodiments the microscope has a
magnification of greater than about 300X. In some embodiments
the microscope has a magnification of greater than about 140X.
In some embodiments the microscope has a magnification of
greater than about 1000X.
[00027] In some embodiments the microscope has a resolution of
about 2.0 to 3.0 microns. In some embodiments the microscope has
a resolution of about 1.5 to 2.0 microns. In some embodiments
the microscope has a resolution of about 1.0 to 1.5 microns. In
some embodiments the microscope has a resolution of less than
about 1.0 microns. In some embodiments the microscope has a
resolution of about 0.88 to 1.0 microns. In some embodiments the
microscope has a resolution of about 0.6 to 0.88 microns. In
some embodiments the microscope has a resolution of 0.2 to 0.5
microns.
[00028] In some embodiments the first stage for supporting the
sample comprises a diagnostic coating for providing a visual
indication to a user through the optical microscope of a
presence of a diagnostic target. In some embodiments the visual
indication is viewed using the optic. In some embodiments the
diagnostic target is a disease, parasite, bacteria, or disorder
detectable in a bodily fluid.
[00029] In some embodiments the optic includes an array of
lenses. In some embodiments the array of lenses simultaneously
shows one or more bright field images and one or more
fluorescence images. In some embodiments the array of lenses
includes four or more lenses.
[00030] In some embodiments the optical microscope is a bright
field microscope. In some embodiments a colored LED and/or light
filter is used and the optical microscope is a fluorescence
microscope, polarization microscope, phase contrast microscope,
etc.
[00031] In some embodiments the optical microscope has bright
field and filter field viewing capabilities wherein a user
shifts from bright field to filter field by lateral movement of
the second stage and the light source that cooperate to provide
either the bright field or the filter field.
[00032] In some embodiments the optical microscope includes a
marking aperture on the second stage configured to allow a user
to identify and indicate a target location on the sample by
marking the first stage through the marking aperture on the
second stage such that a second user can align the optical
microscope to view the target location. In some embodiments, an
optimal aperture is utilized for the series of lenses used in
the microscope in order to minimize the imaging artifacts.
[00033] In some embodiments, the optical microscope includes a
tool for cleaning debris from the lens. The cleaning tool can be
comprised of a glass slide with a piece of lens paper attached
to the surface.
[00034] Methods for using optical microscopes are disclosed
herein. The methods can include engaging a first portion of an
optical microscope configured to support a sample with a second
portion of an optical microscope having a lens; placing a sample
on the first portion of the optical microscope; adjusting the
lens by moving the second portion of the optical microscope
having the lens relative to the first portion of the optical
microscope to focus on the sample; and viewing the sample. In
some embodiments during the adjusting step the optical
microscope has an optical path distance of less than about 3 mm
from the sample to the opposite side of the optic.
[00035] In some embodiments before the adjusting step or the
viewing step is a step of placing a portion of the microscope
against a user's eyebrow. In some embodiments viewing the sample
comprises projecting an image of a portion of the sample. In
some embodiments adjusting the lens includes moving the second
portion relative to the first portion in an out-of- plane
direction from a distance of about 5[mu][iota][eta] to about
625[mu][pi][iota]. In some embodiments viewing the sample
comprises buckling the second portion to adjust the distance
between the lens on the second portion and the sample.
[00036] In some embodiments before viewing the sample is reacted
with a reagent. In some embodiments viewing the sample comprises
testing the sample for a disease.
[00037] In some embodiments the methods include, before
engaging, removing the first portion and the second portion form
a single piece of a flat material. In some embodiments engaging
includes assembling the optical microscope from a flat material.
In some embodiments assembly of the optical microscope involves
origami or folding of a flat sheet of material
[00038] In some embodiments, the sample is reacted to a reagent
already deposited in the sample holding stage via a microfluidic
network embedded in the sample holding stage. In some
embodiments, the reagent is dried for preservation. In some
embodiments, the reagent is wet.
[00039] In some embodiments an optical microscope is provided
including a stage; a lens; and a light source. The optical
microscope having bright field and filter field viewing
capabilities wherein a user shifts from bright field to filter
field by lateral movement of the stage containing a lens and a
light source that cooperate to provide either the bright field
or the filter field. In some embodiments the lateral movement of
the stage to move between bright field and filter field is less
than 10 mm. In some embodiments the lens can be engaged a strip
having tabs with the strip configured to allow a user to slide
the tabs to buckle the strip or adjust the focal length. In some
embodiments sliding the tabs attached to the lens focuses the
image or moves the lens closer to the user's eye.
[00040] In some embodiments an optical microscope is provided,
including a stage; a lens; a light source; and an enclosure
having an enclosed volume of 70 cubic centimeters or less which
contains the stage, lens and light source. In some embodiments
the enclosed volume is 175 millimeters x 70 millimeters x 5.7
millimeters. In some embodiments the optical microscope further
comprises a container configured to store more than twenty
optical microscopes within a volume of 1,400 cubic centimeters
or less. In some embodiments the container configured to store a
range of 20-50 optical microscopes.
[00041] In some embodiments, the lens can be cleaned by
inserting a slide with lens paper attached to the surface and
panning the second stage in circles over the lens paper so that
the lens paper brushes off contaminants from the surface of the
lens. In some embodiments, optical instruments including
interferometers and spectrophotometers are constructed using the
principles disclosed herein to achieve a desired beam path.
BRIEF
DESCRIPTION OF THE DRAWINGS
[00042] The novel features of the invention are set forth with
particularity in the claims that follow. A better understanding
of the features and advantages of the present invention will be
obtained by reference to the following detailed description that
sets forth illustrative
embodiments, in which the principles of the invention are
utilized, and the accompanying drawings of which:
[00043] FIG. 1 illustrates the components for assembly
into an exemplary optical microscope.
[00044] FIG. 2A illustrates a top view of an exemplary
optical microscope.
[00045] FIG. 2B illustrates a cross-sectional view of the
optical microscope of FIG. 2A.
[00046] FIG. 3 illustrates different embodiments of
optical microscopes.
[00047] FIG. 4 illustrates an optical microscope in
accordance with an embodiment.
[00048] FIGS. 5A-E illustrate schematic examples of
optical microscopes at various assembly stages.
[00049] FIGS. 5F-G are various images produced by a
conventional microscope compared to images taken with an
optical microscope in accordance with embodiments.
[00050] FIGS. 6A and 6B illustrate examples of optical
microscopes prior to folding in accordance with various
embodiments.
[00051] FIGS. 7A and 7B illustrate data for the accuracy
and repeatability, respectively, for folding and unfolding an
optical microscope in accordance with embodiments.
[00052] FIG. 8A illustrates an exemplary optical path of
an optical microscope in accordance with an embodiment.
[00053] FIG. 8B is a schematic illustration of an optical
path for an optical microscope in accordance with an
embodiment.
[00054] FIG. 9A illustrates the magnification obtained
versus various lens radii for different refractive index
values.
[00055] FIG. 9B illustrates the third order spherical
aberration versus ball lens radius.
[00056] FIG. 10 illustrates focusing metric versus the
distance between the lens and image for RMS spot size (RSS),
inverse Strehl ratio (1/SR), and ratio of spot size to Strehl
ratio (RSS/SR).
[00057] FIG. 1 1 A illustrates the normalized optical
aperture radius versus magnification for various radii and
refractive index values. FIG. 1 I B illustrates the resolution
versus magnification for various radii and refractive index
values.
[00058] FIGS. 12A-12B illustrate calculations for the
optimal aperture radius and resolution, respectively, as a
function of lens radius, refractive index, and incident light
wavelength.
[00059] FIG. 13 is a graph of resolution versus aperture
radius for various metrics.
[00060] FIG. 14 is a graph of the filter transmission
versus wavelength for various filter types.
[00061] FIGS. 15A and 15B are images of malaria samples
observed with and without a diffuser. [00062] FIG. 16A
illustrate three different LED light sources with and without
a condenser.
FIG. 16B illustrates the intensity of a LED light source.
[00063] FIG. 17A is a model of the ray tracing for a 0.3
mm spherical ball lens.
[00064] FIG. 17B a model of the RMS spot size for
off-axis rays.
[00065] FIG. 17C is a model of an RMS spot radius field
map.
[00066] FIG. 17D is a model of a Strehl Ratio field map.
[00067] FIG. 18 illustrates an image of 1 [mu][eta][iota]
polystyrene beads using a 300 [mu][iota][eta] ball lens and an
aperture size of about 150 [mu][iota][eta].
[00068] FIG. 19A illustrates a bright field image of 1
[mu][iota][eta] polystyrene beads. FIG. 19B illustrates a
fluorescent image of 2 [mu][eta] polystyrene beads. FIG. 19C
illustrates a plot of intensity versus distance. FIG. 19D
illustrates an image of polystyrene beads along with a
2D-Fourier transform showing the power spectrum of the
threshold image and spatial frequency detail.
[00069] FIG. 20A illustrates a schematic for utilizing a
refraction ball lens for projection microscopy with
ray-tracing for half of a lens. FIG. 20B illustrates a
projection illuminated through water containing Fluorescene.
[00070] FIG. 21 A is a projected image of a mosquito
proboscis at an effective magnification of 1500X. FIG. 21 B is
a projected image of red blood cells at an effective
magnification of 3000X.
[00071] FIG. 22A is a picture of a single ball lens
optical microscope submerged under water in accordance with an
embodiment.
[00072] FIG. 22B is a picture of a single ball lens
optical microscope projecting an image on the retina of a user
in accordance with an embodiment.
[00073] FIG. 22C is a picture of a single ball lens
optical microscope projecting an image on a flat surface in
accordance with an embodiment.
[00074] FIG. 23A is an image produced by a single ball
lens optical microscope in accordance with an embodiment of a
ring stage malaria parasite Plasmodium Falciparum in a thin
blood smear stained with Giemsa blue.
[00075] FIG. 23B is an image produced by a single ball
lens optical microscope in accordance with an embodiment of
Trypanosom Cruzi in thin blood smear stained with Giemsa.
[00076] FIG. 23C is an image produced by a single ball
lens optical microscope in accordance with an embodiment of
Giardia stained with Giemsa.
[00077] FIG. 23D is an image produced by a single ball
lens optical microscope in accordance with an embodiment of
sickle-shaped red blood cells.
[00078] FIG. 23E is an image produced by a single ball
lens optical microscope in accordance with an embodiment of
gram positive and gram negative bacteria. [00079] FIG. 23 F is
an image produced by a single ball lens optical microscope in
accordance with an embodiment of Leishmania donovani stained
with Giemsa.
[00080] FIG. 23G is an image produced by a single ball
lens optical microscope in accordance with an embodiment of
Burgia XX stained with Giemsa in a thin blood smear.
[00081] FIG. 23H is a schematic illustration of a 3x3
lens array with different modalities embedded in an optical
microscope in accordance with an embodiment.
[00082] FIG. 231 illustrates a portion of microscopes
with nine (3x3), four (2x2), and 2 parallel optical paths in
accordance with various embodiments.
[00083] FIG. 23 J are bright field images of human blood
cells produced by an optical microscope in accordance with an
embodiment produced by a 3x3 array of lenses
[00084] FIG. 23 are images produced by a single ball lens
optical microscope in accordance with an embodiment of a pine
seed using bright field and polarization microscopy,
respectively.
[00085] FIG. 23L are images produced by a single ball
lens optical microscope in accordance with an embodiment of
multi-fluorescence images of 2 [mu][pi][iota] polychromatic
beads.
[00086] FIG. 24A is an image produced by a single ball
lens optical microscope in accordance with an embodiment of
human chromosomes.
[00087] FIG. 24B is an image produced by a single ball
lens optical microscope in accordance with an embodiment of a
DNA/RNA stain.
[00088] FIG. 24C is an image produced by a single ball
lens optical microscope in accordance with an embodiment of a
spinal cord. &
[00089] FIG. 24D is an image produced by a single ball
lens optical microscope in accordance with an embodiment of
skeletal muscle.
[00090] FIG. 25A is a schematic illustration of a person
stepping on a single ball lens optical microscope. FIG. 25B
shows the undamaged optical microscope after being stepped on.
[00091] FIG. 26 illustrates optical microscopes cut using
a laser in accordance with embodiments.
[
Click tFig. 1 to enlarge ]


[ Click
Fig 4 to enlarge ]












DETAILED DESCRIPTION
[00092] Optical devices and methods for using optical devices,
such as microscopes are disclosed herein. The optical devices
disclosed herein can be assembled from a flat material. The flat
material can include one or more lenses, a sample support,
illumination sources, and control electronics. The optical
devices can be an optical microscope, interferometer, and
spectrophotometers. The optical devices can be designed and
assembled based on Origami principles and optical design
principles. For example, the individual components of the
optical devices can be designed based on Origami principles. The
components can be designed to facilitate assembly and alignment
of the optical device.
[00093] A microscope can be a critical tool for disease
diagnostics, especially in resource poor settings around the
world. Conventional microscopes were designed as general-purpose
research tools. Conventional microscopes are costly and not
designed with portability in mind. Traditional microscopes are
also prone to failure in harsh field conditions since they are
complex three-dimensional objects that are costly, bulky and
require regular maintenance for optimal performance. In
addition, traditional optical instruments require significant
alignment post assembly.
[00094] The optical devices disclosed herein can be assembled in
the field from a flat material and assembled to achieve the
optical alignment. The achievement of the optical alignment
through folding and engagement of the components can be referred
to as passive alignment. Passive alignment can be achieved
through kinematic constraints and structural loops. The optical
devices and individual components of the optical can be shaped
to facilitate passive alignment. Passive alignment great reduces
the costs associated with the optical devices and allows for the
optical devices to be assembled in environments without the
resources for assembling and aligning a conventional optical
device, such as an optical microscope.
[00095] The components of the optical microscope can be formed,
printed, or applied on the flat flexible substrate. Examples of
flat materials include paper, thin metal, polymers, paper coated
with polymers, polyamide, Flex PCBs, etc. The flat material can
be flexible and is not limited to a planar configuration.
[00096] In some embodiments other optical instruments can be
constructed using the folding principles disclosed herein to
achieve a desired optical or beam path in 3 dimensions. Examples
of optical instruments include interferometers,
spectrophotometers, and other optical instruments.
[00097] The optical devices can be assembled from the flat
material using a series of cutting and folding steps. The
optical microscopes disclosed herein can be designed to
self-align due to kinematic constraints present in how the
optical microscope is assembled from the flat materials. The
optical microscope can be designed with folds and mating stages
that self align the one or more lenses with the sample support
and any illumination sources when assembled. Mechanical
components such as a panning and focus stage can be implemented
as flexure mechanisms manufactured by folding paper. The panning
stage can allow for movement in the X-Y direction to position
the optic over any portion of a sample. For a 2 cm by 2 cm
sample the panning stage can include a travel distance of 2 cm
by 2 cm. [00098] The optical device can be a multi-modal
microscope implementing one or more of bright field,
multi-fluorescence, polarization, phase contrast, and projection
microscopy. The optical microscope can include a filter for
fluorescence and polarization modalities. For projection
microscopy the optical microscope includes a light source with a
power suitable to project an image. For example, the light
source may consume 800mA of electrical current at
3.7V.
[00099] The sample can be viewed through the optic. In some
embodiments the user can position the optical microscope close
to their face, such that the optic is located about 5 cm or less
from the user's retina. In some embodiments the optical
microscope can project an image. For example, the image can be
projected on a flat surface such as a screen. Alternatively, the
optical microscope can use a screen designed to conform to the
shape of the Petzval surface of the lens to eliminate field
curvature in the projected image.
[000100] The optical microscope can include an array of lenses.
The lenses can have different sizes and magnifications. The
lenses in the array can also be used to provide images of
different modalities simultaneously, e.g. two different
fluorescence images or a bright field image and a fluorescence
image. The array of lenses can be provided in a side by side
arrangement. The multiple lenses can be paired with one or more
light sources. The array of lenses can be arranged in a grid,
e.g. in a 3x3 grid arrangement.
[000101] The optical microscopes can have an optical path that
is much shorter than conventional microscopes. For example, the
optical path can include the distance between the light source
and the opposing side of the lens. In some embodiments the
distance between the light source and the opposing side of the
optic or lens is from about 1 mm to about 20 mm. In some
embodiments the optical path is less than about 5 mm. In some
embodiments the optical path is less than about 3 mm. The short
optical path allows for the microscope to be a much smaller
vertical height, assuming a vertical optical path, in comparison
to traditional optical microscopes. The shorter vertical height
requires less structural support and less material, thereby
decreasing manufacturing costs. In some embodiments the optical
path between the sample and the opposing side of the optic is
less than about 3 mm. In some embodiments the optical path
between the sample and the opposing side of the optic is less
than about 2 mm or less than about 1 mm. These short optical
paths can allow for illuminating the sample with a a much lower
power light source or lower amount of light versus conventional
microscopes.
[000102] Unlike a traditional microscope, the entire optics and
illumination stage can be panned across the surface of the
sample slide, which can be fixed in a single location. The
optical microscopes disclosed herein can have a lens and light
source that are movable relative to the sample or sample stage.
An optic stage can include a support for the lens. The light
source can be integrated with a separate stage, such as an
illumination stage. The optic stage and illumination stage can
be engaged such that they move together. The lens can move along
above the sample along an x-axis and y-axis to observe different
areas of the sample. In addition the vertical distance, e.g.
along a z-axis, between the lens and sample can be adjusted.
[000103] The optical microscopes disclosed herein can be used
for a variety of applications. The optical microscopes can be
used for diagnostic applications such as testing for a disease,
parasite, bacteria, or disorder that is detectable in a bodily
fluid. Examples include diagnostic applications such as
diagnosing diseases such as Malaria, Chagas, Giardiasis,
Microfilariasis, Sickle-cell disease, and other diseases. Stains
for common parasites and diseases can be used with the samples.
The sample and stain can be observed with the optical microscope
to determine the presence or absence of the diagnostic target.
The optical microscopes can also be used for general purposes,
including science education.
[000104] In some embodiments the lens is a substantially
spherical lens, for example a ball lens. A ball lens
advantageous for manufacturing because it can minimize part
count and be assembled without concern for rotational alignment
as shown in FIG. 8A. Ball lenses can be formed from molten glass
or UV-curable epoxies. Suspension polymerization can also be
used to produce spherical GRIN lenses with reduced spherical
aberration. Other examples of ball lens materials include
borosilicate, sapphire, BK-7, polymeric materials such as
acrylic, etc.
[000105] The ball lens can be sized based on the desired
magnification of the optical microscope. As shown in FIG. 9A the
magnification varies inversely with the ball lens radius. For a
high magnification, e.g. above 500X, the ball lens can have a
small radius, such as below 0.5 mm. The small lens size allows
for a thinner optical microscope thickness and smaller form
factor. In some embodiments the ball lens has a diameter of less
than about 2,500 microns. In some embodiments the spherical ball
lens has a diameter of about 1 ,000 [mu][iota][tau][iota] to
about 2,500 [mu][iota][tau][iota]. In some embodiments the ball
lens has a diameter of less than about 1 ,000 microns. In some
embodiments the spherical ball lens has a diameter of about 300
[mu] [iota] to about 1 ,000 [mu][pi][iota]. In some embodiments
the ball lens has a diameter of about 100 microns to about 1,000
microns. In some embodiments the ball lens has a diameter of
about 100 microns to about 3,000 microns. In some embodiments
the spherical ball lens has a diameter of about 100
[mu][eta][iota] to about 300 [mu][eta][iota].
[000106] In some embodiments half-ball spherical lenses can be
used. In some embodiments a Wallston doublet lens is utilized.
The Wallston doublet lens can be composed of multiple half ball
lenses. In some embodiments a Gradient Index lens is used.
[000107] In some embodiments the effective aperture of the
substantially spherical lens in the optical microscope can be
less than the full diameter of the spherical lens. For example,
a portion of the spherical lens can be covered or removed to
reduce the effective aperture of the spherical lens as discussed
in detail below. In some embodiments the aperture is about 1/4
to about 2/3 of the diameter of the spherical lens. In some
embodiments the spherical ball lens has an aperture diameter of
about 1/4 to about 1/3 of the spherical lens diameter. In some
embodiments the spherical lens has an aperture diameter of about
1/3 to about 1/2 of the spherical lens diameter. In some
embodiments the spherical lens has an aperture diameter of about
1/2 to about 2/3 of the spherical lens diameter.
[000108] The components of the optical microscopes can be
selected to achieve a desired magnification and resolution as
discussed in detail below. In some embodiments the optical
microscope has a magnification of about 100X to about 200X. In
some embodiments the optical microscope has a magnification of
greater than about 140X. In some embodiments the optical
microscope has a magnification of about 200X to about 500X. In
some embodiments the optical microscope has a magnification of
greater than about 300X. In some embodiments the optical
microscope has a magnification of greater than about 500X. In
some embodiments the optical microscope has a magnification of
about 500X to 1 ,500X. In some embodiments the optical
microscope has a magnification of greater than about 1000X. In
some embodiments the optical microscope has a magnification of
greater than about 1500X. In some embodiments the optical
microscope has a magnification of about 1 ,500X to 2,500X. In
some embodiments the optical microscope has a magnification of
greater than about 2000X. In some embodiments the optical
microscope has a magnification of greater than about 2500X.
[000109] In some embodiments the microscope has a resolution of
about 2.0 to about 3.0 microns. In some embodiments the
microscope has a resolution of about 1.5 to about 2.0 microns.
In some embodiments the microscope has a resolution of about 1.0
to about 1.5 microns. In some embodiments the microscope has a
resolution of about 0.88 to about 1.0 microns. In some
embodiments the microscope has a resolution of about 0.6 to
about 0.88 microns. In some embodiments the microscope has a
resolution of less than about 1.0 microns. In some embodiments
the microscope has a resolution of 0.2 to 0.5 microns.
[000110] Examples of light sources include sun light, LED
lights, a lamp, indoor lighting, incandescent lighting,
fluorescent lighting, a flame, chemiluminescence source such as
a glow stick, and other light sources.
[000111] In some embodiments the light source is an LED light.
In some embodiments the LED is white. A white LED can provide a
full-spectrum color image of the sample in bright- field. In
some embodiments the LED is blue. A blue LED can provide light
of an appropriate wavelength for fluorescence imaging. In some
embodiments the LEDs have power suitable so that a user may
perceive multiple simultaneous images of the sample. In some
embodiments the LED has a luminous emittance of about 1 - l
OOkLux or more, which is suitable to project an image of the
sample. In some embodiments the LED has a power of about 10-1
,000 Lux, which is suitable to illuminate the sample so that a
user's eye may perceive the image of the sample.
[000112] In some embodiments the light source can be coupled to
an illumination stage. In some embodiments the optical device or
microscope includes an illumination stage with an element for
modifying the profile of the light from the light source (e.g.
LED), such as a condenser, diffuser, light shaping diffuser,
polycarbonate light shaping filter, etc. The illumination stage
can also include a condenser to focus the light from the light
source. Filters can also be used with the illumination stage.
Examples of filters include polarization filters, polymer color
filters, diffusive filters, and fluorescence filters. In some
embodiments, optical microscope has an illumination stage that
provides Kohler illumination.
[000113] In some embodiments the illumination stage includes an
aperture or hole for allowing light to pass through the
illumination stage towards the sample. The illumination stage
aperture can be sized based on the size and diameter of the
lens. In some embodiments the aperture has a diameter of about
1/4 to about 2/3 of the diameter of the lens
[000114] A power source, such as a battery, can be used to
provide power to the light source. The power source can be
coupled to the illumination stage.
[000115] The sample can be provided to the optical microscope on
a substrate. The substrate can include a coating or reagent. In
some embodiments the substrate can be provided on the flat
material having the optic stage, sample stage, and illumination
stage. In some embodiments the sample is provided on a glass
slide.
[000116] FIG. 1 illustrates the components for assembly into an
exemplary optical microscope. The optical microscope 100
includes sample or specimen stage 1 10, an illumination stage 1
12, a spacer insert 1 14, a locking spacer insert 1 16, and an
optics stage 1 18. In various embodiments, the optical
microscope 100 may not include an illumination stage 1 12. In
the embodiment without an illumination stage 1 12, the
illumination source may be an external source, such as the sun,
a lamp, a candle or some other source. Components 1 10-1 18 may
be part of a single flat sheet of paper that may be used to
implement the optical microscope. The specimen stage 1 10 may
include microscope identifiers 120, grid 122, horizontal
position indicators 124, vertical position indicators 125,
specimen illumination window 126, vertical fine position
indicator 128, horizontal fine position indicator 130, slide
sleeve 132 and slide viewing window 134. In the illustrated
embodiment, horizontal position indicators 124 are denoted by
the letters A-L and vertical position indicators 125 are denoted
by the letters U-Z. Each letter A-L and U-Z of the horizontal
position indicators 124 and vertical position indicators 125,
respectively, may advance in increments of 2mm. Any increment
may be implemented. The microscope identifiers 120 may give the
specification information that may include the microscope
modality and magnification.
[000117] Grid 122 may be used along with horizontal position
indicators 124, vertical position indicators 125, vertical fine
position indicator 128, and horizontal fine position indicator
130. Specimen illumination window 126 may be an aperture or hole
that allows a user to view a sample. Slide sleeve 132 may be
folded over specimen illumination window 126 of specimen stage 1
10. Slide sleeve 132 may include slide slot 136 and slide tab
131. Slide tab 131 may be inserted into tab slot 121 during
assembly. The size of each of the slide sleeve 132, slide slot
136, slide tab 131 , slide slot 136, slide guides 139 and tab
slot 121 may control the force that holds the slide in place. In
various embodiments, the size of each of the slide sleeve 132,
slide slot 136, slide tab 131 , slide slot 136, slide guides 139
and tab slot 121 is configured to allow the slide to be easily
inserted and positioned. In various embodiments, the size of
each of the slide sleeve 132, slide slot 136, slide tab 131 ,
slide slot 136, slide guides 139 and tab slot 121 is configured
to maintain an original position and resist movement after the
sample slide is inserted. A slide containing a sample may be
inserted through slide slot 136 into slide sleeve 132. Slide
guides 139 along the edges of slide sleeve 132 may be folded
over to help guide insertion of the slide. The sample slide (not
shown) may be positioned over specimen illumination window 126.
The edge of the sample slide is lined up so that it is adjacent
to a longitudinal edge 138.
[000118] Specimen stage 1 10 includes a mechanism for
reproducing microscope settings. The mechanism is implemented by
grid 122, vertical fine position indicator 128, horizontal fine
position indicator 130, horizontal position indicator 124,
vertical position indicator 125 and slide sleeve 132.
[000119] The illumination stage 1 12 may include circuitry 140,
light source 142, and contact pads 144A-B. Circuitry 140 may
include copper tape, conductive paper, conductive ink, or any
other flat conductive material. The circuitry 140 can control
the power supply to light source 142. In various embodiments,
the power supply includes a battery. Light source 142 may
include one or more LED lights, a lamp, a chemiluminescence
source such as a glow stick, or other source. Illumination can
be enhanced via a condenser lens 157 (located between 156 and
158). Contact pads 144A-B may establish contact with contacts on
the optics stage 1 18.
Illumination stage 1 12 may be mounted to specimen stage 1 10.
In various embodiments, the optics stage 1 18 and the
illumination stage 1 12 remain aligned within 0.5mm or better
during panning and focusing. This may be accomplished by
connecting tabs 145A-B in the illumination stage 1 12 into the
slots 147A-B in the optics stage 1 18.
[000120] A light source using chemiluminescence uses a chemical
reaction as the illumination or light source. Liquid light
sources may flow into a tube or flat chamber and allow for a
wide array of optical designs. The liquid light source may move
translationally so that the user sees the entire sample.
Usually, one only has 2 degrees of freedom; the present
disclosure allows vision of everything on the slide.
Additionally, the present disclosure may allow for a
configuration of a tube within a tube. A liquid light source
within inner tube may illuminate a sample located between the
inner tube and the outer tube. This design requires "zero power"
to remove the need for a power source such as a battery or
button cells.
[000121] Spacer insert 1 14 and locking spacer insert 1 16 may
be used to position the lens (not shown) and adjust the optics
of the optical microscope 100. One or more spacer inserts 1 14,
1 16 may be attached to illumination stage 1 12. One or more
spacer inserts 1 14 may be positioned first. The locking spacer
insert 1 16 is positioned over the one or more spacer inserts 1
14 to lock or hold them in position.
[000122] Optics stage 1 18 may include contacts 146A-B, vertical
slits 148, horizontal slits 150, and lens apertures 152. Optics
stage 1 18 is mounted to illumination stage 1 12 and specimen
stage 1 10. Contacts 146A-B are designed to align with contact
pads 144A-B. By depressing contacts 146A or 146B, a circuit is
closed with light source 142 and a power source causing light
source 142 to illuminate.
[000123] Vertical slits 148 and horizontal slits 150 align with
vertical fine position indicator 128 and horizontal fine
position indicator 130, respectively. As the sample slide is
moved to the left horizontally, the horizontal fine position
indicators 130 advance in that direction as it goes from one
line to the next. In various embodiments, each horizontal line
of horizontal fine position indicator 130 may represent a
specific increment such as 0.5mm. If a user sees two solid lines
at the same time, the position is in between increments such as
0.25mm. Any increment may be implemented. A location may be
determined by the horizontal position indicators 124 and the
vertical position indicators 125. For example, as a user looks
at the left side of the optical microscope 100, a mark (not
shown) indicates the horizontal position is at A or B. The
vertical position may be determined similarly using vertical
position indicators 125, X, Y, and Z, and may have similar
increments. Vertical fine position indicator 128 and horizontal
fine position indicator 130 may compliment vertical position
indicators 125 and horizontal position indicator 124 by
providing more detailed alignment information.
[000124] The optical microscope 100 may be created using a flat
manufacturing process. Components may be together on a flat
sheet that can be cut out and folded into an optical device. A
design of pre-cut sheets can be created, downloaded, printed and
used. Several integrated microscopes can be contained in one
such optical microscope, creating different optical microscopes
based on how the paper/material is folded. The present
microscope may be a single-use microscope and is suitable for
the present technology because the overall manufacturing process
is inexpensive enough to discard or incinerate after use.
[000125] FIG. 2A illustrates a top view of an exemplary optical
microscope. FIG. 2B illustrates a cross-sectional view of the
optical microscope of FIG. 2A. The optical microscope 100 of
FIGs. 2A-B include specimen stage 1 10, illumination stage 1 12,
optics stage 1 18, and sample 154. Sample 154 may be viewed
through slide viewing window 134 (See FIG. 1). Light source 156
may direct light through filter 158, condenser 157, sample 154,
specimen
illumination window 126 and is received by lens 160. Lens 160
may be positioned between spacer inserts 1 14 and locking spacer
insert 1 16. In some embodiments, a second lens may be
implemented between sample 154 and light source 156. In some
embodiments an array of lenses are used. In some embodiments a
condenser is used to shape the light and improve resolution.
[000126] Focusing the optical microscope 100 can be shown with
reference to FIG. 2B. Focusing the lens 160 can be implemented
by a flexure mechanism that uses a cantilever beam of paper
being pinched at two points for a symmetric upward and downward
motion of the embedded optics as shown in FIG. 2B. Tension or
compression of the optics stage 1 18 causes a Z-axis scan via a
flexure mechanism. The flexure mechanism can have a maximum
travel distance of about 1 mm. The flexure mechanism can
converts purely translational pinching movement along the X-axis
to upward or downward motion along the Z-axis. The distance
between the lens 160 and the sample 154 can be adjusted by
pushing or pulling on the ends of the optics stage 1 18. Pushing
the ends of the optics stage towards each other causes the lens
160 to raise thereby increasing the distance between the lens
160 and the sample 154. Pulling the ends of the optics stage 1
18 away from each other lowers the lens 160 thereby decreasing
the distance between the lens 160 and the sample 154. The
movement of the lens 160 towards and away from the sample 154
can be referred to as movement along the z-axis. In the case
when there are two lenses (as shown in FIG. 5C), the lenses may
be a certain distance apart. In various embodiments, the lenses
may be about 6mm apart. In other embodiments, the lenses may be
about 5.5mm apart. If a user finds something in left lens and
wants to look at the same article/item with the right lens, the
user may put a mark on grid 122. A marking instrument such as a
pen or pencil may be inserted through marking apertures 133 in
the optics stage 1 18 and marking apertures 135 in the
illumination stage 1 12 to mark a position on grid 122. The user
then moves the optics stage 1 18 so that it is lined up with the
mark on grid 122. This may also allow a second user to find and
view the same article/item using markings created by the first
user. For example, a lab technician may be able to easily locate
a particular pathology that has been indicated by markings on
grid 122. The marking apertures can be used in any of the lens
arrangements described herein. The marking apertures can have a
precision of 0.5mm or less.
[000127] The optical design may include focusing features, zoom
features, panning features, and/or other features. Lens 160 may
have a magnification of 140x, 340x, 680x, and 1 140x
corresponding to a borosilicate ball lenses with diameters of
2.4mm, 1.0mm, 0.5mm, and 0.3mm, respectively, as well as other
magnifications. A variety of materials may be used for the lens.
The lens may have a short path length due to the flat optical
design that reduces signal-to- noise ratio for fluorescence. The
path length may be symmetric or asymmetric. An asymmetric path
length is created with a filter, a specimen, another filter, a
ball lens on the top side, a light source and a light receiver.
If a ball lens is placed on the bottom, the microscope becomes
symmetric. The second lens may correct for the LED's divergence,
making the light rays parallel rather than diverging, optimizing
the LED. The LED may also act as a point source depending on the
design.
[000128] FIG. 3 illustrates different embodiments of optical
microscopes with various magnifications and modalities. Bright
field optical microscopes are shown with a magnification of 435X
(1), 1450X (2), and 2175X (3). A fluorescence microscope with
Acridine Orange and Auramine stains is depicted in 5. A
polarization microscope is shown in 6. An optical microscope
with a multi-array of lenses is shown in 7. A projection
microscope is shown in 8. FIG. 4 illustrates an optical
microscope in accordance with an embodiment assembled from a
flat material.
[000129] FIG. 5A shows a schematic arrangement of the parts of
an optical microscope on a flat material that can be cut,
folded, assembled, and used to diagnose a blood sample. FIG. 5B
illustrates various parts of an optical microscope prior to
assembly in accordance with an embodiment. FIGS. 5C, 5D, and 5E
illustrate assembling and adjusting portions of an optical
microscope. FIG. 5C illustrates the x-y scanning feature. FIG.
5D illustrates two lenses 160 with magnifications of 140x and
330x, respectively. The 140x magnification lens 160 produces a
bright field image. The 330x magnification lens 160 produces a
fluorescence image. The glass slide 170 is shown engaged with
the optical microscope 100. FIG. 5E illustrates the z-axis focus
by pushing inward on the edges of the optical stage.
[000130] FIGS. 5F-G are various images produced by a
conventional microscope (FIG. 5F) compared to images taken with
an optical microscope in accordance with embodiments (FIG. 5G).
The images in FIG. 5F were taken using a Nikon Eclipse
fluorescence microscope with a Hg lamp, LED, and glow stick
under a magnification of 400X, respectively. FIG. 5G illustrates
fluorescence images produced by an optical microscope in
accordance with embodiments disclosed herein for AO stained
beads and AO stained white blood cells at a magnification of
330X and 140X, respectively. The images illustrated in FIG. 5F
are comparable to the images achieved using a conventional
microscope shown in FIG. 5G.
[000131] The optical devices disclosed herein can be assembled
without written language instructions. This is advantageous
because the optical microscopes can be used in a variety of
countries without requiring translations of the instructions for
different countries. The language free folding instructions can
include colors and pictures to facilitate the assembly of the
optical microscope. The language free instructions can be
directly printed on the instrument itself, for example a color
matching scheme that guides a user to correct folding sequence.
The surfaces of the optical microscope can also provide staining
instructions for the sample. The surfaces of the optical
microscope can include an identification guide for the
diagnostics provided on the optical microscope. FIGS. 6A and 6B
illustrate examples of optical microscopes prior to folding in
accordance with various embodiments. FIGS. 6A and 6B are
illustrated in black and white but can include bright colors to
facilitate the folding and assembling instructions.
[000132] The optical microscopes can be folded and assembled
with a high degree of precision and accuracy. Traditional
optical instruments require significant alignment post assembly.
An advantageous feature of a folding optical assembly design is
that self-alignment can be exploited with kinematic constraints
present in how the folds are executed, even for manual folding.
Folding a sheet of paper introduces an alignment error
proportional to the thickness of the paper in a valley or
mountain fold. This error arises from uncertain position of a
hinge in a fold due to buckling of paper in the inner edge of
the fold. The alignment errors can be corrected by introducing
folding features that provide kinematic constraints, for example
mating the two stages and implementing closed structural loops
during folding. For a mechanical process, repeatability is
inversely proportional to the square root of contact points for
elastically averaged coupling alignment schemes. The optical
microscopes disclosed herein, for example, can utilize four
contact points and two alignment surfaces to couple the optics
and illumination stage. To characterize the alignment accuracy
and repeatability of the optical microscopes, twenty independent
microscopes were cut out of A4 sheets of paper with a laser
cutting tool having a cutting error of +-6 microns. Each
microscope was hand folded and unfolded repeatedly, while
measuring X-Y alignment errors between the optical and
illumination stage. FIGS. 7A and 7B illustrate data for the
accuracy and repeatability, respectively, for folding and
unfolding the optical microscopes. Assembly accuracy and
precision for the optical alignment as small as 10 microns was
demonstrated for the microscope components in a manual assembly
process using paper with a thickness of about 150
[mu][iota][eta]. The optical alignment can include the alignment
of the light source, optic/lens, sample stage, and any other
items within this path, such as filters, diffusers, apertures,
etc. The alignment precision was acceptable for the lens sizes
disclosed herein, for example a lens having a diameter of 300
[mu][iota][eta].
[000133] The optical microscopes disclosed herein can be durable
and resistant to outdoor elements. For example, the optical
microscopes can be stepped on and still work as shown in FIGS.
25A-B. The microscopes can work submerged in water and survive a
fall from a three story building.
[000134] The optical device can be optimized based on the
desired application or for general use. A number of parameters
can be adjusted to optimize the optical device. For example, the
lens size and shape, lens refractive index, lens material, the
optical path length, optical path shape/direction, the lens
aperture size and shape, the light source type and location, the
aperture associated with the light source, the light source
intensity and profile, the alignment of the light source with
the lens, the wavelength of the light source, the polarization
and coherence of the light source, the magnification of the
sample, the resolution of the image, the inclusion of additional
features such as automated staining or sample separation, the
inclusion of a cell counter feature for keeping track of the
level of parasitimia, optical filters placed in front of the
light source or in front of the lens, etc.
[000135] In some embodiments, the folding accuracy is
accomplished by geometrical features cut in flat material that
act as kinematic couplings thus providing a self-alignment. In
some embodiments, self-alignment is further improved by
providing structural closed loops in folding steps.
[000136] In some embodiments, the optical microscope can have an
integrated microfluidic channel for bringing samples directly to
the microscope lens.
[000137] In some embodiments, the optical microscope can be
incinerated after one use safely and thus can be used with
infected samples. In some embodiments, the entire microscope is
disposable after single or multiple uses.
[000138] In some embodiment, a waveguide is utilized to channel
light from the light source to other optical components.
[000139] In some embodiments, the sample is reacted to a reagent
already deposited in the sample holding stage via a microfluidic
network embedded in the sample holding stage. In some
embodiments, the reagent is dried for preservation. In some
embodiments, the reagent is wet.
[000140] The devices disclosed herein can be modified to modify
the optical path to a desired configuration. The flexibility in
designing the optical path allows the device to be used as a
general purpose optical design tool. Optical devices such as
microscopes, interferometers, and spectrophotometers can be
assembled using the methods, devices, and concepts disclosed
herein. [000141] In some embodiments, the lens can be cleaned by
inserting a slide with lens paper attached to the surface
followed by panning the second stage with the lens/optic in
circles over the lens paper so that the lens paper brushes off
contaminants from the surface of the lens.
[000142] FIG. 8A illustrates an exemplary optical path of an
optical microscope in accordance with an embodiment. FIG. 8A
illustrates the distance between the light source (e.g.LED) and
sample object as D¦ed-0bj, the distance between the sample
object and lens as D0bj-iens, the distance between the lens and
image plane as Dier,s-img- The optical chain of the optical
microscope includes illumination sources (distance Z oi
condenser lens, illumination aperture (A /), sample glass slide,
spherical micro-lens (radius r, refractive index n at a distance
D0bj-iens from the slide) and entrance {A2) and exit {A3)
aperture. For a real image in projection mode, the image plane
is a distance Diem.img apart. The total optical path length from
the light source to the last lens surface can be about 2.5mm,
which is only about 1 % of the optical path length for a
conventional microscope. The decreased optical path length
allows for the microscope to be constructed with a short
vertical height assuming a vertical optical path. The reduced
optical path length can also minimize the extent to which stray
light can enter the system and degrade optical performance.
[000143] FIG. 9A illustrates the magnification obtained versus
various lens radii for different refractive index values. FIG.
9A depicts overall magnification obtained as a function of ball
radius (r) for refractive index (") values ranging from 1.33 to
2. The parameters shown in FIG. 9A can be optimized to improve
the resolution of the optical microscope. Utilizing ray- tracing
methods, a collimated beam of light entering an aperture
diameter A2 and a spherical ball lens with a diameter (D=2r) and
refractive index n, is focused to a point given by the effective
focal length EFL = 2^_^ and numerical aperture (NA) is given by
NA = <A>'^<n> ^ with magnification inversely
proportional to ball lens radius as shown in FIG. 9A. A
magnification of -2175X can be obtained with a sapphire glass
lens of about 200 [mu][eta][iota] in diameter as shown in FIG.
9A. FIG. 9B illustrates the third order spherical aberration
versus ball lens radii. The transverse third order spherical
aberration can be expressed as TSC = ^<71-3>^*^ which
reduces as a
Zn'r"
function of ball lens radius as shown in FIG. 9B.
[000144] The aperture radius (A2) can be optimized to maximize
contrast and resolution, for example by balancing spherical
aberration in the image and transmitted illumination intensity
for a fixed wavelength of light. Considering collimated incident
light, resolution as a function of lens-image distance (D
iens-img) and aperture radius (A2) can be calculated both
numerically and analytically. Since the optimization depends on
two independent parameters numerical optimization is achieved in
two stages. The first stage optimizes the focus by varying D
iens-inig, while the second stage optimizes the resolution by
varying A2 with D iens-img equal to its optimum value for each
aperture radius. For the first optimization stage, the focusing
metric is chosen to be the reciprocal of the Strehl Ratio
(1/SR). Minimizing this focusing metric effectively selects a
value for Diens-img corresponding to diffraction focus, or best
focus. FIG. 10 illustrates focusing metric versus the distance
between the lens and image for RMS spot size (RSS), inverse
Strehl ratio (1/SR), and ratio of spot size to Strehl ratio
(RSS/SR).
[000145] The second optimization stage calculates optimal
aperture diameter (A2) by minimizing a resolution metric,
evaluated as the ratio of Airy Disk Radius and Strehl Ratio
(ADR/SR). Graphically, this metric gives a curve that subtends
the Airy Disk Radius (ADR) for small aperture radii and
parallels RMS Spot Size for large aperture radii. Also, it
provides good numerical convergence. Numerical modeling results
in design curves as a function of lens radius r, refractive
index n, and incident light wavelength [lambda]. FIG. 1 1 A
illustrates the normalized optimal aperture radius versus
magnification for various radii and refractive index values.
FIG. 1 I B illustrates resolution versus magnification for
various radii and refractive index values. FIGS. 12A-12B
illustrate calculations for the optimal aperture radius and
resolution, respectively, as a function of ball lens radius,
refractive index, and incident light wavelength.
[000146] For the analytical model, the optimization proceeds by
evaluating the diffraction limited resolution at best focus,
where the normalized longitudinal aberration ([Lambda]) is equal
to 1. Two different analytical models can be used to derive
identical results, within a multiplicative constant. In the
first analytical model, Airy Disk Radius is equated to RMS Spot
Size
(ADR=RSS). The Airy Disk Radius is evaluated as ADR = 1.22 .
[lambda]- F , where F = EFL/2A2 is the focal ratio or F/#. The
RMS Spot Size is approximated as the RMS blur radius (rRMs)>
at best focus, RSS s
is the focal ratio, s = s- A2<4> is the peak aberration
coefficient and s is the aberration coefficient. For a ball
lens, the following expression was derived for the aberration
coefficient: s = -(" - l)- [" + (2 - ")- (2" - l)]/8rV . Setting
the expression for ADR equal to the expression for RSS and
solving for A2/r, the following expression is obtained for the
normalized optimal aperture radius: [lambda] - [eta]<3>
nOAR = k' , where
- (n - ])- [n + (2 - n)- (2n - \
k[ = (l .22 . 2 . -JeJ = 1.5636 . This can be substituted into
the expression for ADR (or that for RSS) to obtain the following
expression for image resolution:
where k' = (l .227276) 4 = 0.1951. The above expression for
optimal aperture and corresponding resolution show good
agreement with the numerical calculations illustrated in FIGS.
12A-B. The second analytical model determines the
diffraction-limited resolution by minimizing the ratio of Airy
Disk Radius (ADR) to Strehl Ratio (SR), i.e. by solving
d/da(ADR/SR) = 0 . For this model, the Strehl ratio can be
expressed using the empirical approximation,
SR - exp[-(2/w[3/4]/ l)<2>], where [omega]5 is the RMS
wavefront error at best focus. As mentioned earlier, this model
produces the same results as the first model, except with the
constants now being, k[ = (6VK)/7T)<1> = 1.5677 and k2' =
(l .22/4). ([tau][tau]/6)<[iota]/4> . (e/1 o)<1
s>[identical to] 0.2205. For example, the numerical and
analytical models predict an optimized image resolution of about
0.88um for a 300um diameter sapphire lens, which agrees well
with experimental data (see FIG. 12B and FIG. 18). In some
embodiments a resolution can be as high as 500 nm at
magnifications greater than about 2000X.
[000147] The resolution of the optical microscope can be
quantified by observing an object having a known size. For
example, fluorescence resolution was determined between two
adjacent microspheres using ImageJ's linear intensity profile
tool. Microspheres were considered to be resolvable when the
minimum intensity value between the two spheres was 84.4% of the
maximum intensity of the normalized dataset as is in accordance
to the Rayleigh Criterion. To make peaks of intensity
distinguishable, a local polynomial regression model weighted
across every 10 pixels was used to smooth the data. Using this
analysis, the [Iota] [mu][eta][iota] beads were barely
resolvable. FIG. 18 illustrates an image of 1 [mu][pi][iota]
polystyrene beads using a 300 [mu][iota][eta] ball lens and an
aperture size of about 150 [mu][iota][eta]. FIG. 19A illustrates
a bright field image of 1 micron polystyrene beads. FIG. 19B
illustrates a fluorescent image of 2 micron polystyrene beads.
FIG. 19C illustrates a plot of intensity versus distance.
[000148] For brightfield resolution, the beads could clearly be
resolved between one another as shown in FIG. 18. As there are
several methods for measuring resolution in brightfield, we
borrowed a technique used by Lorusso and Joy (2003) for
quantifying resolution using the Fourier Transform of the
bead-matrix image. FIG. 19D is an image of the polystyrene beads
shown in FIG. 18 along with a 2D-Fourier transform showing the
power spectrum of the threshold image and spatial frequency
detail. A threshold was set for the original image such that
noise was removed and the 2-dimensional fast-Fourier transform
was then performed on the image. Upon taking the power spectrum
of the Fourier transform, the extent of the spatial frequencies
was found. The mean radius of this ellipsoidal extent was then
used to calculate the smallest feature size, which is equal to
the resolution, in the image using the following equation:
Resolution = ROIW * PixelSize / RFFT * Image Scaling where ROIW
is the average width of the sensor and RFFT is the average
radius of the power spectrum in the Fourier domain. The pixel
size is the size of the pixels on the CCD. The image scaling is
a factor used in determining the scale of the image plane
compared to the CCD plane. For this image, the ROIW was 4680
pixels, the RFFT was 318.25, the Pixel Size was 6.4 micrometers
and the Scaling factor was 95 CCD pixels per object pixel. This
resulted in a calculated resolution of 0.99 micrometers. In
general, this is a conservative method of measuring resolution
as the smallest features can be filtered out with the noise
during the threshold step.
[000149] FIG. 13 is a graph of resolution versus aperture radius
for various metrics. The optimal resolution metric was obtained
using the ratio of Airy Disk to Strehl ratio. FIG. 14 is a graph
of the filter transmission versus wavelength for various filter
types. FIG. 14 compares an excitation filter, emission filter,
and diffusive filter. Associated fluorescence image data using
the above filter set as barrier (blue) and emission (red) filter
is shown in FIGS. 18 and FIG. 23L. In some embodiments the
filters can be provided with separate light sources. In some
embodiments the filters can be integral with the sample stage.
In some embodiments the filter can be included with the optic
stage.
[000150] Malaria samples are often observed under an oil
immersion lens due to ring artifacts that arise because of
curved shape of a red blood cell. FIGS. 15A and 15B are images
of malaria samples observed with a diffuser. FIG. 15A is an
image obtained using an
embodiment of the optical microscopes described herein and FIG.
15B is an image obtained using a conventional microscope. FIGS.
15A and 15B illustrate that good quality images can be obtained
without the use of an oil immersion lens. Using a simple thin
plastic film diffuser that cuts out directional illumination can
remove artifacts due to cell inhomogeneity and enhance the image
component due to light absorption. An example of the
transmission spectra of the diffusive filter is shown in FIG.
14.
[000151] FIG. 16A illustrate three different LED light sources
with and without a condenser. The three LED based light sources,
labeled #516, #350 and #425, were imaged using fluorescein
dissolved in water. As shown in FIG. 16A, the condenser further
focuses the light. FIG. 16B illustrates the illumination
intensity in a polar plot of LED light source #516.
[000152] Additional modeling is shown in FIGS. 17A-D. A 0.3 mm
sapphire lens is modeled in FIGS. 17A-D. FIG. 17A is model of
the ray tracing for a 0.3 mm spherical ball lens. FIG. 17B a
model of the RMS spot size for off-axis rays of 0, 5, 10, 15,
and 20 degrees. FIG. 17C is a model of an RMS spot radius field
map. FIG. 17D is a model of a Strehl Ratio field map.
[000153] The microscope may also allow a user to view the same
object at two different magnifications at the same time. This
embodiment can provide two concentric magnification regions,
where the inner region is at a higher magnification than the
outer region. This may be achieved by placing a half ball lens
above a smaller diameter ball lens in a doublet configuration,
for example. The microscope may be utilized as a single purpose
microscope, having optics that depend on the type of stain used,
the light source and the filters used. The optics design allows
for illumination to track with the optics. A reflective mylar
sheet may be implemented for providing a mirror surface, and may
allow for simultaneous frontside ackside imaging as shown in
FIG. 8B.
[000154] The present microscope may have a simultaneous
multi-modal imaging feature. Multi-modal may involve two or more
images on the retina simultaneously by taking advantage of the
psychophysics of the eye. The present microscope may have one or
more lenses. Each of the one or more lenses may correspond to a
different microscope modality. The microscope modalities may
include brightfield, darkfield, fluorescence in many sets
depending on the type of filter implemented (GFP, YFP, AO,
Auramine-O, DAPI, or other filter) or any other filter, and
polarization using polarization filters and any other optical
microscopy. Each of the microscope modalities may be implemented
at high and low magnifications. One or more of the microscope
modalities may be implemented at the same time on optical
microscope 100 in any arrangement. For example, the microscope
may include darkfield imaging at a specific magnification and
brightfield imaging at a second magnification. Hence, in the
same field of view, the user may be looking at the same object l
OOOx magnified, and simultaneously looking at it l Ox magnified.
In various embodiments, the microscope modalities form an array.
[000155] In various embodiments, the ball lenses may be arranged
in an 3x3, 4x4 or 5x5 matrix, where each lens provides identical
images but with a slightly different experience - a continuous
field may be presented. For example, if looking at red blood
cells, even though each ball is only contributing one portion of
that field, it appears to be like a continuous field. The images
provided by the matrix merge as one and appear as a landscape
rather than individual images. In various embodiments, if a user
changes an illumination angle, and turns it bright or dark, the
microscope may provide the user with 3D information about the
sample by producing contrast over the sample based on variations
in the sample's 3D profile. Because the illumination stage can
move, a user can take the same object and achieve a 3D imaging
effect.
[000156] Another feature of the optical microscope 100 is a
projection mechanism. FIG. 20A illustrates a schematic for
utilizing a refraction ball lens for projection microscopy with
ray- tracing for half of a lens. FIG. 20B illustrates a
projection illuminated through water containing fluorescein. The
projection mechanism can project images onto a wall, screen, or
other surface. The mechanism is a brighter LED such as a
high-intensity LED to make the projection more visible. Rather
than looking directly into the optical microscope, a user may
flip it over and point to a wall in the dark. Two or more people
may be looking at the same image at the same time. For example,
in a dark room, a high-intensity LED may allow a projection of
an image from across the room, making a 1 mm mosquito proboscis
appear over 2 meters tall on the wall. The projection may be
onto a camera, a wall, a retina, or other location, and the
microscope can be a digital image of the projection - the
projection optics are the same. In various embodiments, there is
also a CCD version that may be provided with one or more cameras
integrated in the present system.
[000157] The projected mode allows for a number of people to
simultaneously view the images. The effective magnification in
projection mode is dependent on the distance between the lens
and the projected image with the intensity inversely
proportional to the distance between the lens and projected
image. FIG. 21 A is a projected image of a mosquito proboscis at
an effective magnification of 1500X. FIG. 21B is a projected
image of red blood cells at an effective magnification of 3000X.
[000158] FIG. 22B is a picture of a single ball lens optical
microscope projecting an image on the retina of a user in
accordance with an embodiment.
[000159] Optical components of the manufacturing process may
include microfluidics, doublets, and other features.
Microfluidics may be used to manufacture components. Doublets
may be created from liquid and cured lenses and used in
combination with a microscope. This may allow for lower costs
and timing while increasing magnification. A unique slide
position readout may provide a location grid of the sample and
allow a user to know where they are on the slide. An integrated
cell count mechanism may also be used with the present optical
microscope.
[000160] Fluidic integration into microscope design may minimize
sample handling. An auto-staining process within the microscope
involves filters designed from plastic films. The auto-staining
process may have the staining dye already integrated.
[000161] Several optical microscopes may be contained in a
single package. The optical microscopes can be packaged
individually similar to a band aid and revealed just before use.
This packaging minimizes the risk of having fungus grow on
microscope components in humid environments.
[000162] The optical microscopes disclosed herein can also be
used under water. The microscope can include a water resistant
coating such as a polymer coating. Operating under water allows
for live imaging of microscopic swimming organisms under water.
The underwater image can be projected on the side wall of the
aquarium. The thermal gradients generated allow for water to
flow past by the sample stage, thus bringing live swimming
organisms to be imaged in projection mode through the
microscope. FIG. 22A is a picture of a single ball lens optical
microscope submerged under water in accordance with an
embodiment.
[000163] The optical microscopes disclosed herein can be used
for disease diagnostics. For example, temperature stable stains
are widely available for labeling infectious disease samples
including Giemsa and Acridine Orange that are usually used on a
standard thin blood smear. Various imaging parameters of optical
microscopes disclosed herein, such as field of view,
magnification, etc. can be optimized to match the imaging
requirements for a disease of interest. The optical microscopes
disclosed herein can be configured and built to match a specific
disease of interest. In some embodiments disease specific
optical microscopes are used instead of general-purpose optical
microscopes. For example, various optical microscope
configurations were setup and utilized to image freshly prepared
diagnostics samples of various parasitic diseases, including
Plasmodium falciparum (FIG. 23A), Trypanosom cruzi (FIG. 23B),
Giradia lamblia (FIG. 23C), sickle cell disease (FIG. 23D), gram
positive and gram negative bacteria (FIG. 23E), Leishmania
donovani (FIG. 23F) and Dirofilaria immitis (FIG. 23G) imaged
via various magnification settings. It is noted that the above
data (maximum magnification 2500X) was obtained without the use
of oil immersion technique or coverslips, which further
complicates sample preparation and microscope maintenance in
conventional diagnostics assays. This broad dataset reveals the
versatility of utility of the optical microscopes disclosed
herein in disease diagnostics.
[000164] FIG. 24A is an image produced by a single ball lens
optical microscope in accordance with an embodiment of human
chromosomes. FIG. 24B is an image produced by a single ball lens
optical microscope in accordance with an embodiment of a DNA/RNA
stain. FIG. 24C is an image produced by a single ball lens
optical microscope in accordance with an embodiment of a spinal
cord. FIG. 24D is an image produced by a single ball lens
optical microscope in accordance with an embodiment of skeletal
muscle.
[000165] Microscopy based diagnostics can be further improved
via integrated multiple imaging modalities (such as bright field
and fluorescence imaging) to improve sensitivity or increase the
total field of view via implementation of parallel array
microscopes. The optical microscopes disclosed herein can be
packed in a close form to implement Simultaneous Multi- modal
Microscopy (SMM) using the miniature optical components and
independent optical paths disclosed herein. FIG. 23H is a
schematic illustration of a 3x3 lens array with different
modalities embedded in an optical microscope in accordance with
an embodiment. The use of SMM is largely possible due to the
large surface area of the retina (FIG. 23H) that allows for many
fields of view to be packed in a closed packed configuration,
such as 1 +1 , 2x2 or 3x3 array microscopes (FIG. 231) with nine
fields of view visible simultaneously (FIG. 23J). Since the
optical path for each field of view is independent, different
modes of microscopy can be combined to build simultaneous bright
and polarization setup (FIG. 23K) or fluorescence from two
different wavelengths (FIG. 23 L). Simultaneous Multi-modal
Microscopy (SMM) enables imaging of samples in multiple
modalities simultaneously, which is not possible in conventional
optical setups.
[000166] FIG. 8B is a schematic illustration of an optical path
for an optical microscope in accordance with an embodiment. FIG.
8B illustrates two ball lenses 160. One ball lens 160 directly
observes the sample 154 using the light source 142. The eye can
directly observe the sample 154 through the optical path 201 and
ball lens 160. Simultaneously, the eye can observe the sample
through an alternate optical path 203 involving the second ball
lens 160 and a mirror205 that can be tilted to focus on the back
of the sample. The mirror can be adjusted to observe a desired
area of the sample 154. In some embodiments multiple lenses can
be aligned with one or more mirrors to alternately view the
sample.